CHN Analyzer
Categories: Mining Engineering Lab
About this product
- Field Application – Coal / Coke and other carbonaceous rock material
- Conforming to the international standard ASTM D 5373.
- Carbon, Hydrogen, Nitrogen must be detected using Infrared / Thermal Conductivity Detector.
- Instruments must have the facility for internal calibration by analyzing standard sample.
- Balance – having accuracy of 0.1 mg. with automatic weight entry facility to analyzer/equipment.
- Instrument should be capable of analyzing sample portion of 50 mg to 200 mg of 212 micron size coal/coke sample.
- Power to the instruments – 230 ± 10% V AC, 50Hz.
- System must have fully automatic operation facility at user command basis. Auto sampler should have 30 or more positions.
- Tender submitted must carry supportive documents for the above specifications and range of detection of above elements repeatability/accuracy as per international standard etc., with branded computer ( Dell, HP, Compaq, Lenovo etc.) having following specifications- Processor i7 or better, 500 GBHDD, 4 GB RAM, 19” LED colour monitor with USB ports, multimedia keyboard, optical mouse and LaserJet printer.
- Supplier should provide at least one certified reference coal sample with certified values of C, H, N.
- Gas, Cylinders & Regulators – Equipment should have the capability to use Helium gas as carrier gas and Oxygen as combustion gas. Two gas filled (He & O2) 47 liter water capacity steel cylinder and two double stage SS regulators (BIS certified) should be supplied with equipment (Price should be quoted separately).
- Tender submitted must carry the copy of recent 2 to 3, purchase order for similar instrument.
- Warranty – 5 years comprehensive for complete instruments
- Consumable for testing of 5000 sample must be supplied with the instrument. It should be quoted separately.
- Others – Installation, training and demonstration on samples for a minimum period of five days or till satisfaction whichever is greater.
Related Products
Mohs Hardness Kit
Mohs Hardness Kit Specification:- The set includes eight points - for Mohs hardness of 2,...
Turbid Meter
Specification:- Turbidity meter bench microprocessor controlled automatic calibration,...
Dean & Stark Unit
Specification:- Nominal voltage 110 V - 230 V AC; Frequency 50 / 60 Hz; Nominal wat...
Electro Analyzer
Specification:- Laboratory apparatus for determinations of lead, nickel, copper, zinc,...
Thermal Conductivity Unit
Specification:- Measuring principle transient heat flow (needle probe method) M...
Loviband Refractometer
Loviband Refractometer ManufacturersSpecification:- The Lovibond Brix Refractometer ran...
Soxhlet Unit
Specification:- Soxhlet Extraction Unit Mantle type heating system. Ideal heater for...
Strohelin Unit
Specification:- Strohlein (carbon &sulphur apparatus) semi-automatic apparatus is u...